Blood flow velocity vector field reconstruction from dual-beam bidirectional Doppler OCT measurements in retinal veins

Gerold C. Aschinger, Leopold Schmetterer, Veronika Doblhoff-Dier, Rainer A. Leitgeb, Gerhard Garhöfer, Martin Gröschl, René M. Werkmeister*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

9 Citations (Scopus)

Abstract

In this paper, we demonstrate the possibility to reconstruct the actual blood flow velocity vector field in retinal microvessels from dual-beam bidirectional Doppler optical coherence tomography measurements. First, for a better understanding of measured phase patterns, several flow situations were simulated on the basis of the known dual beam measurement geometry. We were able to extract the vector field parameters that determine the measured phase pattern, allowing for the development of an algorithm to reconstruct the velocity vector field from measured phase data. In a next step, measurements were performed at a straight vessel section and at a venous convergence; the obtained phase data were evaluated by means of the new approach. For the straight vessel section, the reconstructed flow velocity vector field yielded a parabolic flow. For the venous convergence, however, the reconstructed vector field deviated from a parabolic profile, but was in very good accordance with the simulated vector field for the given vessel geometry. The proposed algorithm allows predictions of the velocity vector field. Moreover, the algorithm is also sensitive to directional changes of the flow velocity as small as <1°, thereby offering insight in the flow characteristics of the non-Newtonian fluid blood in microvessels.

Original languageEnglish
Article numberA005
Pages (from-to)1599-1615
Number of pages17
JournalBiomedical Optics Express
Volume6
Issue number5
DOIs
Publication statusPublished - Apr 6 2015
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2015 Optical Society of America.

ASJC Scopus Subject Areas

  • Biotechnology
  • Atomic and Molecular Physics, and Optics

Fingerprint

Dive into the research topics of 'Blood flow velocity vector field reconstruction from dual-beam bidirectional Doppler OCT measurements in retinal veins'. Together they form a unique fingerprint.

Cite this